    Calibration of sealed HCl cells used for TCCON instrumental line shape monitoring

    Thumbnail
    View/Open
    Hase(2013).pdf (6.559Mb)
    Authors
    Hase, F.
    Drouin, B.J.
    Roehl, C.M.
    Toon, G.C.
    Wennberg, P.O.
    Wunch, D.
    Blumenstock, T.
    Desmet, F.
    Feist, D.G.
    Heikkinen, P.
    De Maziere, M.
    Rettinger, M.
    Robinson, J.
    Schneider, M.
    Sherlock, V.
    Sussmann, R.
    Te, Y.
    Warneke, T.
    Weinzierl, C.
    Show allShow less
    Discipline
    Earth and related Environmental sciences
    Subject
    calibration
    error analysis
    FTIR spectroscopy
    greenhouse gas
    monitoring
    network analysis
    quality control
    Audience
    Scientific
    Date
    2013
    Metadata
    Show full item record
    Description
    The TCCON (Total Carbon Column Observing Network) FTIR (Fourier transform infrared) network provides highly accurate observations of greenhouse gas column-averaged dry-air mole fractions. As an important component of TCCON quality assurance, sealed cells filled with approximately 5 mbar of HCl are used for instrumental line shape (ILS) monitoring at all TCCON sites. Here, we introduce a calibration procedure for the HCl cells which employs a refillable, pressure-monitored reference cell filled with C2H2. Using this method, we identify variations of HCl purity between the TCCON cells as a non-negligible disturbance. The new calibration procedure introduced here assigns effective pressure values to each individual cell to account for additional broadening of the HCl lines. This approach will improve the consistency of the network by significantly reducing possible station-to-station biases due to inconsistent ILS results from different HCl cells. We demonstrate that the proposed method is accurate enough to turn the ILS uncertainty into an error source of secondary importance from the viewpoint of network consistency.
    Citation
    Hase, F.; Drouin, B.J.; Roehl, C.M.; Toon, G.C.; Wennberg, P.O.; Wunch, D.; Blumenstock, T.; Desmet, F.; Feist, D.G.; Heikkinen, P.; De Maziere, M.; Rettinger, M.; Robinson, J.; Schneider, M.; Sherlock, V.; Sussmann, R.; Te, Y.; Warneke, T.; Weinzierl, C. (2013). Calibration of sealed HCl cells used for TCCON instrumental line shape monitoring. , Atmospheric Measurement Techniques, Vol. 6, Issue 12, 3527-3537, DOI: 10.5194/amt-6-3527-2013.
    Identifiers
    uri: https://orfeo.belnet.be/handle/internal/2913
    doi: http://dx.doi.org/10.5194/amt-6-3527-2013
    scopus: 2-s2.0-84890483029
    Type
    Article
    Peer-Review
    Yes
    Language
    eng
    Links
    NewsHelpdeskBELSPO OA Policy

    Browse

    All of ORFEOCommunities & CollectionsBy Issue DateAuthorsTitlesDisciplinesThis CollectionBy Issue DateAuthorsTitlesDisciplines
     

    DSpace software copyright © 2002-2016  DuraSpace
    Send Feedback | Cookie Information
    Theme by 
    Atmire NV