    Calibration results obtained with Liulin-4 type dosimeters

    View/Open
    Dachev(2002a).pdf (677.8Kb)
    Authors
    Dachev, T.
    Tomov, B.
    Matviichuk, Y.
    Dimitrov, P.
    Lemaire, J.
    Gregoire, G.
    Cyamukungu, M.
    Schmitz, H.
    Fujitaka, K.
    Uchihori, Y.
    Kitamura, H.
    Reitz, G.
    Beaujean, R.
    Petrov, V.
    Shurshakov, V.
    Benghin, V.
    Spurny, F.
    Show allShow less
    Discipline
    Physical sciences
    Subject
    Calibration
    Dosimeters
    Monte Carlo methods
    Protons
    Radiation
    Spectrometers
    Angular sensitivities
    Space research
    heavy ion
    pollution exposure
    radiation budget
    aerospace medicine
    aircraft
    article
    calibration
    comparative study
    cyclotron
    equipment design
    instrumentation
    magnetic and electromagnetic equipment
    Monte Carlo method
    neutron
    radiation dose
    radiation monitoring
    radiometry
    theoretical model
    Aerospace Medicine
    Aircraft
    Calibration
    Cyclotrons
    Equipment Design
    Heavy Ions
    Models, Theoretical
    Monte Carlo Method
    Neutrons
    Particle Accelerators
    Radiation Dosage
    Radiation Monitoring
    Radiometry
    Audience
    Scientific
    Date
    2002
    Metadata
    Show full item record
    Description
    The Mobile Radiation Exposure Control System's (Liulin-4 type) main purpose is to monitor simultaneously the doses and fluxes at 4 independent places. It can also be used for personnel dosimetry. The system consists of 4 battery-operated 256-channel dosimeters-spectrometers. We describe results obtained during the calibrations of the spectrometers at the Cyclotron facilities of the University of Louvain, Belgium and of the National Institute of Radiological Sciences-STA, Chiba, Japan with protons of energies up to 70 MeV. The angular sensitivities of the devices are studied and compared with Monte-Carlo predictions. We also present the results obtained at the HIMAC accelerator with 500 MeV/u Fe ions and at the CERN high energy radiation reference fields. Records made during airplane flights are shown and compared with the predictions of the CARI-6 model.
    Citation
    Dachev, T.; Tomov, B.; Matviichuk, Y.; Dimitrov, P.; Lemaire, J.; Gregoire, G.; Cyamukungu, M.; Schmitz, H.; Fujitaka, K.; Uchihori, Y.; Kitamura, H.; Reitz, G.; Beaujean, R.; Petrov, V.; Shurshakov, V.; Benghin, V.; Spurny, F. (2002). Calibration results obtained with Liulin-4 type dosimeters. , Advances in Space Research, Vol. 30, Issue 4, 917-925, DOI: 10.1016/S0273-1177(02)00411-8.
    Identifiers
    uri: https://orfeo.belnet.be/handle/internal/5180
    doi: http://dx.doi.org/10.1016/S0273-1177(02)00411-8
    scopus: 2-s2.0-6444245918
    Type
    Article
    Peer-Review
    Yes
    Language
    eng
    Links
    NewsHelpdeskBELSPO OA Policy

    Browse

    All of ORFEOCommunities & CollectionsBy Issue DateAuthorsTitlesDisciplinesThis CollectionBy Issue DateAuthorsTitlesDisciplines
     

    DSpace software copyright © 2002-2016  DuraSpace
    Send Feedback | Cookie Information
    Theme by 
    Atmire NV